Pulse Oximeter
Alarms
56
System alarm messages:
Message
Source
SYST ALARM: bat low!
The battery voltage is below a level that guarantees
reliable operation.
SYST ALARM: MS-FIFO!
An error has arisen in the Masimo input data memory.
SYST ALARM: MS-timeout!
The Masimo-circuit board has not responded to several
queries.
SYST ALARM: MS-comm.!
The Masimo-data transfer has a permanent defect.
SYST ALARM: MS-error!
The Masimo circuit board has reported an irreparable
defect.
SYST ALARM: Unexp.reset!
An unexpected device restart has occurred.
SYST ALARM: NVRAM error!
The internal non-volatile memory has reported an error.
SYST ALARM: WDT error!
The integrated circuit that monitors the has
reported an error.
SYST ALARM: Stack overfl.
An error has occurred in the software stack processing.
SYST ALARM: ROM-CRC!
A data security error has occurred.
Defective loud speaker!
The has detected a defective alarm loudspeaker.
Содержание sat 801+