7.5 Alarm limits (
Alarm limits
)
CAUTION!
Changing the alarm settings is a serious intervention in the device functions and
must never be done without first consulting the doctor in charge of the treatment!
NOTICE!
Until the modified alarm limit value has been accepted, the most recently stored value
remains in force.
If the alarm limit modification procedure is interrupted or cancelled, this value persists.
If implausible alarm limits are set (e.g. the
lower limit value is greater than the higher
value), the device responds with the error
message „Alarm limits are not correct“ to
the Accept request.
In this case, the alarm limits selected are
not accepted as new values.
NOTICE!
The difference between the upper
and lower alarm limit values must, to
be plausible, be at least two units.
Fig. 29: Alarm limits not correct
7.5.1 Setting the alarm limits
1.
Starting from the normal display, call:
Alarm limits
Fig. 30: Alarm limits