Pulse Oximeter
Technical data
13
NOTICE!
A functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a
pulse oximeter
.
NOTICE!
If there is independent demonstration that a particular calibration curve is accurate for the
combination of a pulse oximeter and a pulse oximeter probe, then a functional tester can
measure the contribution of a monitor to the total error of a monitor/probe system. The
functional tester can then measure how accurately a particular pulse oximeter is
reproducing that calibration curve
.
Perfusion
Display range
0.02-20.0 %
Setting range of the alarm limits
Upper Limit oxygen saturation
42-100 %
Lower Limit oxygen saturation
40-98 %
Upper Limit pulse frequency
27-240 1/min
Lower Limit pulse frequency
25-238 1/min
Alarm tone loudness
75 dbA
Sensor
Wavelengths
660 nm (red), 905 nm (infrared)
max. light intensity
0.79 mW
Notice!
Information about wavelength range can be especially useful to clinicians.