BT-780 Operation manual

 

47 

P/N : 780-ENG-OPM-EXP-R00 

Bistos Co., Ltd. 

2020.03 

Module 

Option 

Module defaults 

Adult 

Pediatric 

Neonate 

TEMP 

Alarm level 

Mid 

Mid 

Mid 

Alarm record 

Off 

Off 

Off 

Display color 

White 

White 

White 

Temperature unit 

T1 alarm limits 

36.0~39.0 on 

36.0~39.0 on 

36.0~39.0 on 

T2 alarm limits 

36.0~39.0 on 

36.0~39.0 on 

36.0~39.0 on 

TD alarm limits 

0.0~2.0 on 

0.0~2.0 on 

0.0~2.0 on  

21 Common faults and maintenance 

The following table shows the common faults on the operation, and the solution. 

Faults 

Solution 

Blank Screen 

Connects the monitor to check the screen and screen line whether normal. 

The system time is not correct 

1. Set up error, can be reset through the system User Maintenance

menu.

2. The button battery on main control board is run out, please change

the button battery.

No ECG waveform 

1

See the ECG cable and lead-wires whether in good condition,

disconnected or electrode rusting result in connection fail.

2

 

Look at whether the ECG cable and lead type are consistent. 

Unable to do ST analysis 

1

Check the ECG Setup

 ST Analysis

 ST Analysis is set to “On”.

2

 

Check the ECG Setup

 Other Setup

 Paced whether be set to “On”.

If Paced is set to “On”, means the patient have a pacemaker; in this 

case the machine is not doing the ST analysis. 

No SpO

2

 waveform or value 

Check whether the SPO2 Sensor is connected and in good condition. 

Blood pressure does not start 

1

Check whether the pump is broken.

2

 

Check whether the trachea is broken.

3

 

Check whether the blood pressure plate is normal.

Blood pressure started, but couldn't 

measure the value 

1

Check whether the blood pressure cuff is leakage.

2

 

Check whether the NIBP extension tube and machine connect is well.

3

 

Check whether the deflating valve on blood pressure plate is normal.

4

 

Check whether the pressure sensor is normal.

If the above doesn't solve the problem, please contact Bistos after-sales department or dealers.

 

22 Manufacturer’s declaration on EMC 

BT-780 needs special precautions regarding EMC (Electromagnetic compatibility) and needs to be used according to the EMC 

information provided in this user manual. Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect the BT-780 and should be kept at least 1 m away from 

the equipment. 

NOTE

 

 

Using  unqualified accessories, sensors and cables will increase the electromagnetic emission and reduce the

electromagnetic immunity of the device.

 

Do not put the device close to other devices or stack together. When necessary, observe the device closely to

ensure that it runs normally in the environment.

 

The device requires special EMC protection, and it is necessary to install and maintain it in the environment that

meets the following EMC information.

 

Even if other devices comply with CISPR emission requirements, they may also cause interference to this device.

 

When the input signal amplitude is smaller than the minimum amplitude specified in the technical specifications,

it may result in inaccurate measurements.

 

Mobile communication devices or wireless network devices may have an impact on the device.

22.1 Electromagnetic emissions 

The BT-780 is intended for use in the electromagnetic environment specified below.  

The customer or the user of the BT-780 should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions 

CISPR 11 

Group 1 

The BT-780 uses RF energy only for its internal function. Therefore, its RF 

emissions are very low and are not likely to cause any interference in 

nearby electronic equipment. 

RF emissions 

CISPR 11 

Class A 

The BT-780 is suitable for use in all establishments other than domestic, 

Содержание BT-780

Страница 1: ...BT 780 Patient monitor Operation Manual Keep this manual for future reference P N 780 ENG OPM EXP R00...

Страница 2: ...r no responsibility is assumed by Bistos for its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implication or otherwise und...

Страница 3: ...ng waves 16 3 7 Other common setup 16 4 Patient information management 17 4 1 Patient setup menu 17 4 2 Admitting a patient 17 4 3 Patient information 18 4 4 Discharging a patient 18 4 5 Clear alarms...

Страница 4: ...Review 35 13 1 Reviewing trend chart 36 13 2 Reviewing trend table 36 13 3 NIBP measurement review 36 14 Battery 37 14 1 Overview 37 14 2 Battery usage guide 37 14 3 Checking battery performance 37 14...

Страница 5: ...nown this information thoroughly before using BT 780 IPX1 Indicates the protection level against the ingress of liquid IPX1 is protection against some water drops falling vertically It correspond the...

Страница 6: ...val is once per year If the hospital or healthcare institutions using this device fail to implement a satisfactory maintenance schedule it will result in device failure and may endanger the patient s...

Страница 7: ...f the person or organization who makes an unauthorized modification or incorporates an unapproved attachment to the device An operator may only perform maintenance procedures specifically described in...

Страница 8: ...mental temperature exceeds 60 Do not heat or splash the battery or throw it into fire or water Do not destroy the battery Do not pierce battery with a sharp object such as a needle Do not hit with a h...

Страница 9: ...tory illness e g asthma chronic obstructive pulmonary disease Monitoring during administration of respiratory depressant drugs e g opiate epidural or patient controlled analgesia Assessing oxygen satu...

Страница 10: ...ture recordings To enable close observation in resolving hypothermia hyperthermia To observe and monitor patients for changes indicating an infection To monitor the effect of treatment for antimicrobi...

Страница 11: ...e1 2 Side view Figure1 3 Rear view 1 5 Description of monitor Figure1 4 Front view Name Description 1 Alarm indicator Indicates the priority of physiological alarm and technical alarms in different co...

Страница 12: ...Rear view Name Description 1 Air outlet Heat dissipation 2 Handle Handle for main body transport 3 Speaker holes For alarm and synchronizing sound 4 ID label Identify the monitor information 5 USB po...

Страница 13: ...f each parameter module The parameters are shown in fixed position that is from top to bottom and from left to right ECG NIBP SpO2 and PR TEMP RESP 4 Information Tip Area Display the network status ba...

Страница 14: ...n any organization or individual shall not interpolate copy or exchange by any means or form When the monitor is combined with other devices it must comply with IEC 60601 1 2005 A1 2012 and should not...

Страница 15: ...let If using battery power make sure the battery is fully charged Check all the functions required for patient monitoring to make sure that the monitor operates properly WARNING If the monitor is dama...

Страница 16: ...enus is basically similar to the Settings mode as shown below You can set up the monitor screen if you need Menu title A title of the current menu Close menu Close the current menu Exit the current me...

Страница 17: ...e screen Select Department Enter the sector and department using the device through the soft keyboard on the screen Select Bed Number Enter the bed number through the soft keyboard on the screen 3 7 2...

Страница 18: ...y is very important If the monitor has admitted the patient it is recommended to operate the monitor to discharge the current patient before connecting to not admitted the next patient Otherwise the d...

Страница 19: ...onate 7 Select Pace Maker and set whether the patient wears a pacemaker 8 Select Height cm and set the patient s height via the pop up keyboard on the screen Range 0 250 9 Select Weight kg and set the...

Страница 20: ...the display format according to needs and get different screen information 5 1 Selecting user interface Select the user interface as follows Select Displays smart hotkey Screen Select Select the displ...

Страница 21: ...s the alarm limit or the patient has physiological abnormalities The information of physiological alarm is displayed in the physiological alarm area on top of the screen Technical alarms Technical ala...

Страница 22: ...ority alarm Beep beep beep Low priority alarm Beep 6 3 3 Alarm information Alarm information displayed on the physiological or technical alarm area of the monitor indicates the corresponding alarm inf...

Страница 23: ...utes 6 4 4 Alarm off As shown in 6 4 3 if the Alarm Pause Time is set to Permanent press the Pause smart hotkey and the monitor will turn off the alarm In this case except the alarm prompt characteris...

Страница 24: ...tive range of the system and the upper alarm limit must be higher than the lower alarm limit Select SpO2 Low Limit and set the lower limit of SpO2 alarm Select SpO2 High Limit and set the upper limit...

Страница 25: ...for defibrillation proof so the monitor operates normally after defibrillation NOTE In the factory setup ECG wave display in the first two waves from top in the wave area in the normal display format...

Страница 26: ...isual alarm 7 3 2 Selecting lead Select the ECG parameter area or wave area ECG Setup menu Select Other Setup ECG Other Setup menu Select Lead Type and select the ECG lead as needed 3 Lead 3 lead ECG...

Страница 27: ...art hotkey and select Patient Info menu Select Yes No for Pace Maker as needed indicating that the patient with or without pacemaker Diagnostic Monitor Surgery will not affect rejection of pacemaker p...

Страница 28: ...be set Auto 0 25 0 5 1 2 4 Select Wave Speed and set the wave speed The wave speed is 12 5mm s 25mm s 50mm s The default is 25mm s Select Filter and set the filter mode Diagnostic Diagnostic mode Mon...

Страница 29: ...ion electrode placement 8 3 1 Adjusting position of respiration electrode If you want to measure ECG and respiration simultaneously you may need to adjust the position of the two electrodes for respir...

Страница 30: ...ty In RESP Setup menu select sensitivity and set the sensitivity The sensitivity is 1 2 3 4 5 The default is 2 8 6 Alarm setup Select Alarm Setup Alarm Setup interface to set respiration related alarm...

Страница 31: ...will have deviation When the patient has a tendency to hypoxia use the oximeter to analyze blood samples in order to fully grasp the patient s condition Do not put the sensor on limbs with arterial du...

Страница 32: ...ncy radio wave interference such as interference generated by the host system or interference from electrosurgery instrument connected to the system Intravenous dye Too frequent movement of the patien...

Страница 33: ...fter complete a measurement start the next measurement continue 5 minutes then stop During measurement any error will stop the continuous measurement If the first measurement time is over 4 minutes an...

Страница 34: ...beats min the blood pressure measurement is impossible Obese patients A thick layer of fat around a limb blocks the arterial oscillation so that it cannot reach the cuff The accuracy is lower than no...

Страница 35: ...tient is neonate the pressure can be select from 100 120 The default cuff pressure value is 100 11 7 NIBP reset Select NIBP parameter area NIBP Setup menu Select Reset and restore the inflation pressu...

Страница 36: ...mperature in the specified environment temperature range or else it may be dangerous 12 5 Temperature display The monitor can display the body temperature of two channels T1 and T2 and the alarm limit...

Страница 37: ...r values of the time which will change automatically with the move of trend cursor 13 2 Reviewing trend table Select Trend smart hotkey to enter Trend menu select Tabular and enter the following windo...

Страница 38: ...ry out of the reach of children Use only the designated battery If the battery is damaged or leaks replace it immediately Do not use a defective battery for the monitor Do not open the battery compart...

Страница 39: ...manufacturer s instructions Do not use abrasive materials such as steel wool Do not allow any liquid entering into the enclosure and never immerse any part of the device into liquid Do not leave any c...

Страница 40: ...ue is not accurate NIBP leakage test At least once every two years or follow hospital regulations NIBP calibration At least once every two years or follow hospital regulations 16 4 ECG calibration In...

Страница 41: ...In accordance with classification specified in the European Medical Device Directive 93 42 EEC this monitor is Class IIb device The monitor is classified as follows in accordance with IEC 60601 1 Cate...

Страница 42: ...mm mV x0 5 10mm mV x1 0 20mm mV x2 0 40mm mV x4 0 Wave sweep speed 12 5 mm s 25 mm s 50 mm s Band width Diagnostic mode 0 05 100 Hz Monitor mode 0 5 40 Hz Surgery mode 1 25 Hz CMRR 100 dB Notch 50 60...

Страница 43: ...Resolution 0 1 Measurement accuracy 0 3 Number of channel 2 Alarm T1 T2 upper limit 0 1 50 0 0 1 step T1 T2 lower limit 0 49 9 0 1 step TD upper limit 0 50 0 0 1 step RESP Measurement method Thoracic...

Страница 44: ...ring with arterial blood sample reference measured with a CO oximeter Pulse oximeter measurements are statistically distributed and about two thirds of the measurements are expected to come within the...

Страница 45: ...orks abnormally check the pacemaker Pacemaker Not work H ECG Signal weak H The patient ECG signal is too weak and the system can t analyze Check the patient s condition electrodes cables and leads ST...

Страница 46: ...s Manual is used Check the sensor mounting position whether the sensor is damaged or sensor type Reconnect the sensor or use new sensor SpO2 Sensor Off L SpO2 Search Timeout L SpO2 Search Pulse L Sens...

Страница 47: ...settings On On On On Mid Mid Mid Mid Off Off Off Off Gain x1 x1 x1 Wave velocity 25 0mm s 25 0mm s 25 0mm s Filter mode Monitor Monitor Monitor Wave color Green Green Green Wave style Color scale Col...

Страница 48: ...whether the pressure sensor is normal If the above doesn t solve the problem please contact Bistos after sales department or dealers 22 Manufacturer s declaration on EMC BT 780 needs special precauti...

Страница 49: ...and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Страница 50: ...is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transm...

Страница 51: ...g BT 780 This product is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on...

Отзывы: