BT-780 Operation manual
35
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
calculate the difference between the body temperature data.
The monitor is designed for defibrillation proof, so the monitor operates normally after defibrillation.
12.2 Safety information
WARNING
Before monitoring, check if the probe cable is normal. Unplug the temperature probe cable from the jack, the
screen will display “TEMP1 Sensor Off” and “TEMP2 Sensor Off” prompt and make an alarm sound.
Calibrate the temperature measuring instrument at least once every two years (or according to hospital
procedures). When calibration is required, please contact Bistos.
12.3 Measurement steps
Please refer to the following steps:
1.
Turn on the monitor and check if it works normally.
2.
Select the appropriate temperature probe according to the patient category and measurement needs.
3.
Insert the probe lead wire into the temperature probe interface.
4.
Attach the probe to the patient properly.
5.
Make sure that the alarm settings apply to the patient.
When measuring body temperature, temperature probe can be attached to body surface such as the neck, armpits, ears and other
locations.
12.4 Measurement requirements
The normal measuring range of monitor is 0
℃
~50
℃
, and the accuracy is consistent in this range.
The environmental temperature range for body temperature measuring is 5
℃
~40
℃
. Get the right temperatures for the shortest
measurement time is 40s, and the measuring interval is 1s.
WARNING
Please measure the body temperature in the specified environment temperature range, or else it may be
dangerous.
12.5 Temperature display
The monitor can display the body temperature of two channels (T1 and T2) and the alarm limits, difference between the two
temperature (TD) and temperature units. Select Temp parameter area and open the [Temp Setup] menu.
Temperature display area is as shown below:
Figure 12-1: TEMP parameter display
WARNING
The operator, prior to use, need to check the compatibility of the probe and thermometer. If the temperature value
displayed by the monitor has significant difference from the body temperature under normal condition, please
check if the probe resistance of the monitor matches the resistance set in the monitor system; if not, please replace
a probe with appropriate resistance or adjust the monitor and select the appropriate resistance. Incompatible
probe will affect the critical properties.
12.6 Setting temperature unit
You can define your favorite temperature unit as follows:
Select TEMP parameter area
→
“TEMP Setup” menu.
In the “TEMP Setup” menu, set “Unit” to “
℃
” or “
℉
”
12.7 Alarm setup
In “TEMP Setup” menu, select “Alarm Setup >>” to enter “Alarm Setup” interface, and set TEMP alarm switch, alarm level, upper
and lower alarm limit. See 6.5 Alarm Setup for detailed setting method.
12.8 Technical description
Accessories have passed the biocompatibility test and meet the requirements of ISO 10993-1.
13 Review
The monitor provides up to 168 hours trend data review of all monitoring parameters, 1000 groups of NIBP measurement data and
200 physiological alarm events, 100 technical alarm events. The user can select trend chart or trend table to view trend change; or
view the latest wave.