BT-780 Operation manual
41
P/N : 780-ENG-OPM-EXP-R00
Bistos Co., Ltd.
2020.03
Display
Type
Color TFT touch screen LCD
Size and resolution
15.6”, 1366*768 pixels
Audio
Speaker
Alarm sound (45 ~ 85 dB), key pressing sound
QRS sound, PR sound
Alarm sound meet the IEC 60601-1-8 standard requirements
Alarm signal
Alarm delay
Off, 1s, 2s, 3s, 4s, 5s, 6s, 7s, 8s, depending on the setup
Default is 4s
Pause duration
1min, 2min, 3min, 4min, 5min, 10min, 15min or permanent, depending on the
setup
Default is 2 minutes.
Data storage
Trend
168 hours. Resolution: 1 min
Alarm event
200 physiological alarm events, 100 technical alarm events
NIBP measurement result
1000 groups
Environment
Operation
Transport and storage
Temperature
5~ 40°C (41°F~104°F)
–20 ~ 60°C (-4°F~140°F)
Humidity
30~ 85% non-condensing
0 ~ 95 % non-condensing
Atmospheric pressure
70~106 kPa
70~106 kPa
18.3 Functional specifications
18.3.1 ECG/TEMP/RESP
ECG
Standard compliance
IEC 60601-2-27:2012
Lead Type
5 lead
I, II, III, aVR, aVL, aVF, V
12 lead (optional)
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Display sensitivity
Auto, 2.5mm/mV(x0.25), 5 mm/mV(x0.5),
10mm/mV(x1.0), 20mm/mV(x2.0), 40mm/mV(x4.0)
Wave sweep speed
12.5 mm/s, 25 mm/s, 50 mm/s
Band width
Diagnostic mode
0.05 - 100 Hz
Monitor mode
0.5 - 40 Hz
Surgery mode
1 - 25 Hz
CMRR
>100 dB
Notch
50/60 Hz notch filter can be set to on or off
Differential input impedance
> 5 M
Ω
Electrode polarization voltage
range
± 400 mV
Baseline recovery time
<5s after defibrillation (in monitor and surgery mode)
Calibration signal
1 mV (peak
–
peak), accuracy
±
3%
Lead-off detection current
Measuring electrode: < 0.1 uA
Drive electrode: < 1uA
Pacing pulse
Pulse identification
For PACE MAKER pulses that meet the criteria below, pacing pulse will be marked
on the screen.
Detection range(Amplitude): ± 2 mV ~ ± 700 mV
Pulse width: 0.2ms ~ 2.0 ms
Average HR
Calculate from 15s data
Interval of HR refreshing
Calculate once every second
HR change response time
Time from 80 bpm to 120 bpm:
≤
10 sec
Time from 80 bpm to 40 bpm:
≤
10 sec
Tall T-wave suppression
For T-wave with 100ms QRS wave, 350ms QT period, 180ms duration and 1.2mV
amplitude, the HR calculation will not be affected
Without overshoot rejection of
pacemaker pulses
Amplitudes (ap) from ±2 mV to ±700 mV and pulse widths from 0.1 ms to 2.0 ms.
Tall T-wave rejection capability
2mV
HR
Measuring range
Adult: 15 ~ 300 bpm
Pediatric/Neonate: 15 ~ 350 bpm