background image

BT-700 Operator Manual                                             15 

 

P/N : 700-ENG-OPM-EUR-R02 

Bistos Co.,Ltd. 

2018,03 

 

Fig. 4.2 Main Monitoring Screen – Numeric Mode 

 

 

Fig. 4.3 Main Monitoring Screen – Spo2 Trend Mode 

 

 

Fig. 4.4 Main Monitoring Screen – Pulse rate Trend Mode 

Содержание BT-700

Страница 1: ...BT 700 Patient Monitor OPERATOR MANUAL Keep this manual for future reference P N 700 ENG OPM EUR R02 BT 700 ...

Страница 2: ...tos Corporation is believed to be accurate and reliable However no responsibility is assumed by Bistos for its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implication or otherwise under any patent or patent rights of Bistos Revision 02 Mar 2018 Copyright Bistos Corporation 2011 All rights reserved 7th FL A Bldg Woolim L...

Страница 3: ...nection 13 4 BT 700 OPERATION 14 4 1 Factory Setting 14 4 2 BT 700 Display Screen 14 4 2 1 Parameter Frame 16 4 2 2 SpO2 Frame 16 4 2 3 Pulse Rate Frame 16 4 2 4 Alarm limits Frame 16 4 2 5 Plethysmograph frame 16 4 2 6 Temperature frame 16 4 2 7 SpO2 trend frame and Pulse rate trend frame 16 4 2 8 Alarm message Time frame 17 4 2 7 Status frame 17 4 3 BT 700 Controls and Indicators 17 4 4 BT 700 C...

Страница 4: ... EUR R02 Bistos Co Ltd 2018 03 7 CLEANING AND DISINFECTION 23 7 1 Monitor 23 7 2 Sensors 23 8 TROUBLESHOOTING AND MAINTENANCE 25 8 1 Sensors 25 8 2 Battery Disposal and Handling 25 8 3 Maintenance 25 8 4 Troubleshooting 25 9 SPECIFICATION 26 ...

Страница 5: ...operation during the use of defibrillators or during defibrillator discharge The BT 700 is not specified or intended for operation in the presence of electrosurgical equipment The BT 700 is not specified or intended for operation in conjunction with any other type of monitoring equipment except the specific devices that have been identified for use in this Operator s Manual Perform periodic safety...

Страница 6: ...cer gel and other substances CAUTION The relevant law restricts this device to sale by or on the order of a physician CAUTION When installing the unit into a cabinet allow for adequate ventilation accessibility for servicing and room for adequate visualization and operation CAUTION Do not operate the unit if it is damp or wet because of condensation or spills Avoid using the equipment immediately ...

Страница 7: ...cessories Follow cleaning and disinfection instructions in Section 7 of this manual CAUTION Do not immerse the SpO2 sensor in liquid When using solutions use sterile wipes to avoid pouring fluids directly Follow cleaning and disinfection instructions in Section 7 of this manual CAUTION If the equipment use in area where the integrity of the external protective conductor in the installation or its ...

Страница 8: ...om 10 C to 40 C Operating humidity ranges from 30 to 85 Avoid in the vicinity of Electric heater Avoid placing in an area where there is an excessive humidity riseor ventilationproblem Avoid placing in an area where thereisanexcessive shockorvibration Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage Avoid dust and especially metal material into the equ...

Страница 9: ...8 This symbol identifies a safety note Ensure you understand the function of this control before using it Control function is described in the operator s manual IEC 60601 1 External Signal IN OUT Port IEC TR 60878 Defibrillation proof type BF applied part IEC60601 1 IPX7 IPX7 Water proof Temperature type rectal IEC60529 IPX2 IPX2 Water proof SpO2 sensor IEC60529 ...

Страница 10: ...t amounts of dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin In contrast hemoximeters such as the IL482 report fractional saturation oxygenated hemoglobin expressed as a percentage of all measured hemoglobin including measured dysfunctional hemoglobins To compare functional saturation measurements to those from an instrument that measures fractional saturation fractional measur...

Страница 11: ...ional If you wish to purchase please contact our office Disposable SpO2 sensor probe Temperature sensor probe Type Rectal I V holder WARNING The Temperature sensor of BT 700 is used only rectal type The Temperature sensor of BT 700 is calibrated with YSI temperature sensor WARNING BT 700 and The SpO2 sensor are certificated with international standards The SpO2 sensor of BT 700 is type nellcor SpO...

Страница 12: ...r Indicating LED AC Green Battery Orange TFT Color LCD Mode Change Button Trend Save Button Alarm On Off Button Bright Button Control Knob Enter then Menu and Control Beep Volume Mode Change Status Led Trend Save Status Button Alarm On Off Button 3 2 Description of the Side Panel Fig 3 2 Left Panel Fig 3 3 Right Panel SpO2 Sensor Connector SpO2 Sensor Holder Temperature Sensor Connector ...

Страница 13: ...r Technical engineer USB Port Connector For Firm Ware Upgrade Only for Technical engineer Dip Switch Speaker SpO2 Sensor Holder 3 4 Power On Probe Connection When the user want to turn BT 700 on power adaptor is connected with BT 700 below Figure 3 5 and power button is pressed for 2 seconds refer to Fig 3 1 Also SpO2 Sensor probe is connected below Figure 3 5 Fig 3 5 Power adaptor and Probe Conne...

Страница 14: ...e unaffected when the monitor is powered down A complete list of these parameters is shown below Setting Parameter Factory Default Monitoring Mode Graph mode Target Adult Alarms Volume 3 ID 00000 Name Empty Data Interval 1 minute SpO2 Limits High Adult 100 Neonate 80 SpO2 Limits Low Adult 85 Neonate 80 PR Limits High Adult 170 bpm Neonate 190 bpm PR Limits Low Adult 40bpm Neonate 90 bpm Alarm Peri...

Страница 15: ...Operator Manual 15 P N 700 ENG OPM EUR R02 Bistos Co Ltd 2018 03 Fig 4 2 Main Monitoring Screen Numeric Mode Fig 4 3 Main Monitoring Screen Spo2 Trend Mode Fig 4 4 Main Monitoring Screen Pulse rate Trend Mode ...

Страница 16: ...percentage of hemoglobin binding sites in the bloodstream occupied by oxygen 4 2 3 Pulse Rate Frame The Pulse Rate Frame displays beats per minute of the Patient s pulse rate 4 2 4 Alarm Limits Frame The Alarm Limits Frame displays the Alarm Limits set by operator 4 2 5 Plethysmograph Frame The Plethysmograph Frame displays a graphical representation of the blood perfusion 4 2 6 Temperature Frame ...

Страница 17: ...suring patient is Adult Neonate Icon Indicates that the measuring patient is Neonate Battery Status Icon Indicates the battery charge status 4 3 BT 700 Controls and Indicators There are seven buttons located on the front panel The buttons are activated by pushing with the finger until an audible click is heard The operation of the buttons is summarized below Symbol Description Turn power on or off...

Страница 18: ...has been selected for editing the knob is rotated to scan through the available choices for this parameter Pressing the knob stores the new value temporarily 4 5 Menu Tree The Menu Tree of BT 700 is such as table 4 4 Level 1 Level 2 Level 3 Default setting System Setup Target Adult Language English Alarms SpO2 High Low limits Adult H 100 L 85 Neonate H 95 L 80 PR High Low limits BPM Adult H 170 L ...

Страница 19: ...end to start saving When the function is stared the save icon is activated and rotated Press mode button Trend again to stop saving 4 6 2 Trend Data Data list Control the knob to enter the saved data list of the Trend data In data list the saved data is patient ID Save start time and end time Select ID by controlling Knob Figure 4 x is displayed Fig 4 4 Trend Mode display Select the saved list by ...

Страница 20: ...pply If the artifact pulse rate indication ceases the monitor may be used normally If these measures do not result in cessation of the pulse rate artifact the monitor cannot be safely used in this environment 5 2 Basic operation The BT 700 uses pulse oximetry to measure functional oxygen saturation in the blood Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed such a...

Страница 21: ...display to determine whether the battery has sufficient charge to complete the monitoring session Use the AC power if the battery is too low Step 2 Check the sensors the SpO2 sensor and the temperature sensor Check the SpO2 sensor or the Temperature sensor to verify proper attachment to the monitor Check the emitter red light and the photo detector are properly aligned If measure the SpO2 and the ...

Страница 22: ...finger Sensor error Faulty SpO2 sensor is connected Medium Priority exceeding the SpO2 Limits exceeding the Pulse rate Limits detached sensor to the BT 700 Low Priority low battery 6 2 Alarms silence period Alarms can be silenced for preset period called alarm silence duration To view the current setting press ALARM button until 30sec 60 sec 120sec or off is displayed To adjust the setting enter M...

Страница 23: ...lean and disinfect only according to the following instructions 1 Wipe the device with a sterile wipe soaked in enzymatic detergent safe for use with metal instruments Wipe the exterior of the device three times Prepare the detergent according to the manufacturer s sensor recommendations 2 Scrub sensors with enzymatic detergent using soft bristled brush for five 5 minutes DO NOT IMMERSE SENSORS WA...

Страница 24: ...use including all warnings cautions and instructions WARNING Do not use a damaged sensor or pulse oximetry cable Do not use a sensor with exposure optical components WARNING Use only one pulse oximetry cable to increase the length of the sensor Use of more than one pulse oximetry cable may have an adverse effect on performance Do not attach any cable that is intended for computer use to the sensor...

Страница 25: ...battery Adhere to all applicable laws regarding recycling Avoid storing battery above 60 C 140 F If clothing or skin comes in contact with material from inside the battery immediately wash with plenty of clean water When you replace internal Ni MH contact to our technical support office 8 3 Maintenance BT 700 requires periodic calibration or adjustment The recommended interval for performing hipot...

Страница 26: ...on Defibrillation proof type BF applied part Probe Water proof Level SpO2 sensor IPX2 Rectal type temperature sensor IPX7 EMC Class Group I Class B Power Internal 6 V 1 2 V 5ea Ni MH rechargeable 4 hours Fast Charge Continuously 6 hours External AC DC Adaptor Input AC100 240 V 50 60 Hz Output DC 9 V 1 5 A Power Consumption 80VA maximum SpO2 sensor Wavelength RED 660 nm IRED 940nm Environmental Ope...

Страница 27: ...50 BPM Perfusion Range 0 3 20 SpO2 and Pulse rate Accuracy and Motion Tolerance SpO2 Without motion Adult 2 digits of range Without motion Neonate 3 digits of range With motion Adult and Neonate 3 digits of range Pulse Rate Without motion Adult 2 digits of range Without motion Neonate 3 digits of range With motion Adult and Neonate 3 digits of range Temperature Range and Accuracy Range and accurac...

Страница 28: ...od 2 Years Probe excluded Date of Purchase Customer Hospital Address Name Telephone Sales Agency Manufacture BISTOS Co Ltd Thank you for purchasing BT 700 This product is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Consumer s protection law noticed by Economic Planning Dept ...

Страница 29: ... Telephone 82 31 750 0340 Fax 82 31 750 0344 BISTOS Co Ltd 7th FL A Bldg Woolim Lions Valley 5 cha 302 Galmachi ro Jungwon gu Seongnam si Gyeonggi do Korea www bistos co kr bistos bistos co kr Model Name BT 700 Obelis s a Bd Général Wahis 53 1030 Brussels BELGIUM Telephone 32 2 732 59 54 Fax 32 2 732 60 03 ...

Отзывы: