BT-700 Operator Manual 5
P/N : 700-ENG-OPM-EUR-R02
Bistos Co.,Ltd.
2018,03
Section
1
Safety
1.1 Instructions for the Safe Operation and Use of the
BT-700
Examine the monitor and any accessories periodically to ensure that
the cables, line cords, transducers, and instruments do not have visible
evidence of damage that may affect patient safety or monitoring
performance. The recommended inspection interval is once per week
or less. Do not use the monitor if there is any visible sign of damage.
Only the AC line cord supplied with the BT-700, or its equivalent, is
approved for use with the Unit.
Do not attempt to service the BT-700 monitor. Only qualified service
personnel should attempt any needed internal servicing.
The BT-700 is not specified or intended for operation during the use of
defibrillators or during defibrillator discharge.
The BT-700 is not specified or intended for operation in the presence
of electrosurgical equipment.
The BT-700 is not specified or intended for operation in conjunction
with any other type of monitoring equipment except the specific
devices that have been identified for use in this Operator’s Manual.
Perform periodic safety testing to insure proper patient safety. This
should include leakage current measurement and insulation testing.
The recommended testing interval is once per year.
1.2
Warnings
WARNING
: EXPLOSION HAZARD — Do not use the BT-700 in a
flammable atmosphere where concentrations of flammable anesthetics or
other materials may occur.
WARNING
: Be informed that it may cause serious injury or death to the
patient, property damage, material losses against the “Warning" sign.
CAUTION
: Be informed that it may cause no harm in life but lead to
injury against the“ Caution" sign.