Bird 4421 RF Power Meter
30
General Bus Commands
The general bus commands supported by the RS-232 interface module
are listed in Figure 25.
IN
i
T
ialize
(INT)
Function
Resets the Bird 4421 and returns it to the factory defaults.
Remarks
If INT is linked with any other command within a string, it must be
separated from that command by a space.
ENT
er
(ENT)
Function
Makes the power meter transmit a reading to the controller.
Remarks
A measurement must have already been triggered, placing a reading
in the output buffer.
To send a reading whenever a measurement is triggered, set DIP
switch 2 to OFF. The ENT command will not need to be sent.
TR
i
G
ger
(TRG)
Function
Initiates a measurement if the power meter is in trigger mode (T3).
Figure 25
RS-232
General Bus
Commands
Command
Effect on Bird 4421
INT
Returns to default conditions
ENT
Sends a reading to the controller
TRG
Triggers reading in T3 mode
B1 to B7
Selects a baud rate
XO/XF
Enables/disables software handshake