IEEE-488 GPIB Interface
21
R
ange
(Rxx)
Function
Selects a measurement range listed in Figure 16.
Remarks
If the selected range is outside the range of the connected power
sensor, the command is ignored.
Terminators
(Yx)
Function
Selects the characters that follow the end of a data string. Set x to:
y
“T” for two terminators; a carriage return(CR) and a line feed(LF).
y
“O” for one terminator; a carriage return(CR).
y
“N” for no terminator; message can be terminated by EOI.
Remarks
Many controllers use the terminator sequence to recognize the end of
an input string. Using incorrect terminators can lock the bus.
Figure 16
Measurement
Ranges
Command
Power Range
RYY
Turn auto range on
RNN
Turn auto range off. Keep present range
R17
18.0 – 199.9 MW
R16
1.80 – 19.99 MW
R15
0.180 – 1.999 MW
R14
18.0 – 199.9 kW
R13
1.80 – 19.99 kW
R12
0.180 – 1.999 kW
R11
18.0 – 199.9 W
R10
1.80 – 19.99 W
R09
0.180 – 1.999 W
R08
18.0 – 199.9 mW
R07
1.80 – 19.99 mW
R06
0.180 – 1.999 mW
R05
18.0 – 199.9 µW
R04
1.80 – 19.99 µW
R03
0.180 – 1.999 µW
R02
18.0 – 199.9 nW
R01
1.80 – 19.99 nW
R00
0.180 – 1.999 nW
Содержание THRULINE 4021
Страница 2: ...This page intentionally left blank...
Страница 14: ...Bird 4421 RF Power Meter 4...
Страница 18: ...Bird 4421 RF Power Meter 8...
Страница 48: ...Bird 4421 RF Power Meter 38...