RS-232 Interface
31
B
aud Select
(Bx)
Function
Selects a baud rate listed in Figure 26.
Remarks
When the meter recognizes a valid Bx command, its baud rate is
immediately changed. (The controller is assumed to be transmitting at
the new rate; otherwise sending commands would not be possible.)
This command overrides the DIP switch setting.
X
mission Flow Control
(XO/XF)
Function
Enables or disables the XON/XOFF flow control.
y
XO enables flow control.
y
XF disables flow control.
Remarks
When data is being sent from the power meter to the computer and
flow control is enabled, data transmission will be suspended when the
XOFF character (hexadecimal 13) is sent by the computer.
Transmission will resume when XON (hex 11) is sent by the computer.
When data is being sent from the computer to the power meter, XOFF
will be sent to the computer when the input buffer fills up. XON will
be sent to the computer when the buffer has emptied.
Figure 26
Baud Rates
Command
Baud Rate
B1
110
B2
300
B3
600
B4
1200
B5
2400
B6
4800
B7
9600