Bird 4421 RF Power Meter
46
Bird 4020 Series RF Power Sensors
Power Range
4021, 4022
4024, 4025
300 mW – 1 kW
3 W – 10 kW
Frequency Range
4021
4022
4024
4025
1.8 – 32 MHz
25 MHz – 1 GHz
1.5 – 32 MHz
100 kHz – 2.5 MHz
Accuracy, Fwd
± 3%
Insertion VSWR (Insertion Loss)
4021, 4024, 4025
4022
1.05 (0.05 dB) max
1.05 (0.05 dB) max, 25 – 512 MHz
1.10 (0.13 dB) max, 512 MHz – 1 GHz
Directivity, Minimum
4021, 4022
4024
4025
30 dB
30dB (2.5 – 25 MHz), 28 dB (1.5 – 32 MHz)
30 dB (125 – 2500 kHz), 28 dB (100 – 125 kHz)
Signal Purity
For rated accuracy no more than 1% AM
Harmonics –50 dB or less
Содержание THRULINE 4021