Chapter 4
Operating Instructions
Operation without Interlock
Normal Operation
1. Turn on the coolant supply.
NOTE: Minimum flow rate is 4 gal/min (15 L/min) at 5 °C (40 °F).
The minimum flow increases linearly up to 6 gal/min (23 L/min) at
60 °C (140 °F.)
2. Wait for the coolant flow to stabilize.
3. Apply RF power.
Shutdown
1. Turn off RF power at the source.
2. Wait at least five minutes for the Econoload to cool.
3. Turn off the water supply.
Always
do this last.
Emergency
Shutdown
Turn off RF power at the source.
CAUTION
Do not interrupt coolant supply. Even momentary
application of RF power while coolant is not circulating could
cause immediate destruction of the load.