40
Parameters
Pacing and Sensing SR(-T)
Pulse amplitude
and pulse width
Sensitivity
Ventricular capture control
Lead configuration
IEGM recordings
Rates for statistics
Parameter
Range of values
Standard
Pulse amplitude
0.2 ... (0.1) ... 6.2; 7.5 V
3.0 V
Pulse width
0.1; 0.2; 0.3; 0.4; 0.5; 0.75; 1.0; 1.25;
1.5 ms
0.4 ms
Parameter
Range of values
Standard
Sensitivity
AUTO
0.5 ... (0.5) ... 7.5 mV
AUTO
Parameter
Range of values
Standard
Ventricular capture control
ON; ATM (monitoring only)
OFF
ON
Min. amplitude
0.7 V
0.7 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
3.0 V
Safety margin
0.3 ... (1) ... 1.2 V
0.5 V
Search time
Interval; time of day
Interval
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
Time of day
00:00 ... (15 min) ... 23:45 hh:mm
02:00 hh:mm
Parameter
Range of values
Standard
Pacing polarity
Unipolar; bipolar
Unipolar
Sensing polarity
Unipolar; bipolar
Unipolar
Parameter
Range of values
IEGM recordings
20 (quantity); each max. 10 s
Types of IEGM recordings
High rate (HR)
Patient triggered (Pt.)
IEGM recording prior to event
0; 25; 50; 75; 100%
IEGM signal
Filtered; unfiltered
Parameter
Range of values
Standard
High rate (HR)
150 ... (5) ... 200 bpm
400; 387 ... 308; 300 ms
180 bpm
333 ms
HF counter
4; 8; 12; 16
8