38
Parameters
Pacing and Sensing DR(-T)
Pulse amplitude
and pulse width
Sensitivity
Atrial capture control
Ventricular capture control
Parameter
Range of values
Standard
Pulse amplitude A
0.2 ... (0.1) ... 6.2; 7.5 V
3.0
Pulse width A
0.4; 0.5; 0.7; 1.0; 1.2; 1.5 ms
0.4 ms
Pulse amplitude V
0.2 ... (0.1) ... 6.2; 7.5 V
3.0 V
Pulse width V
0.4; 0.5; 0.7; 1.0; 1.2; 1.5 ms
0.4 ms
Parameter
Range of values
Standard
Sensitivity A
AUTO
0.1 ... (0.1) ... 1.5 ... (0.5) ... 7.5 mV
AUTO
Sensitivity V
AUTO
0.5 ... (0.5) ... 7.5 mV
AUTO
Parameter
Range of values
Standard
Atrial capture control
ATM (monitoring only); OFF
OFF
Min. amplitude
0.5 ... (1) ... 4.8 V
1.0 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
3.0 V
Safety margin
0.5 ... (1) ... 1.2 V
1.0 V
Search time
Interval; time of day
Interval
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
Time of day
00:00 ... (15 min) ... 23:45
02:00
Parameter
Range of values
Standard
Ventricular capture control
ON; ATM (monitoring only); OFF
ON
Min. amplitude
0.7 V
0.7 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
3.0 V
Safety margin
0.3 ... (1) ... 1.2 V
0.5 V
Search time
Interval; time of day
Interval
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
Time of day
00:00 ... (15 min) ... 23:45 hh:mm
02:00 hh:mm