35
Parameters
Ventricular
pacing suppression
Upper rate
Mode switching
Refractory periods
Parameter
Range of values
Standard
V
p
suppression
OFF; ON
OFF
Pacing suppression after
consecutive V
s
1 ... (1) ... 8
6
Pacing supports after
X-out-of-8 cycles
1; 2; 3; 4
3
Parameter
Range of values
Standard
Upper rate
90 ... (10) ... 200 ppm
130 ppm
Atrial upper rate
OFF;
240 ppm
240 ppm
Parameter
Range of values
Standard
Mode switching
OFF; ON
ON
Intervention rate
100 ... (10) ... 250 ppm
160 ppm
Switch to (mode)
DDI;
DDI(R) when permanent DDD(R)
VDI;
VDI(R) when permanent VDD(R)
DDIR
VDIR
Onset criterion
3 ... (1) ... 8
5
Resolution criterion
3 ... (1) ... 8
5
Change of the basic rate
with mode switching
OFF;
+5 ... (5) ... +30 ppm
+10 ppm
Rate stabilization with
mode switching
OFF; ON
OFF
Parameter
Range of values
Standard
Atrial refractory period
AUTO
AUTO
Atrial refractory period in
the modes AAI(R); AAT(R);
DDT
300 ... (25) ... 775 ms
350 ms
PVARP
AUTO;
175 ... (5) ... 600 ms
AUTO
PVARP after PVC
PVARP + 150 ms (max: 600 ms)
is automatically programmed
400 ms
Ventricular
refractory period
200 ... (25) ... 500 ms
250 ms