20
Prior to Implantation
Preparing the Implantation
Have parts ready
• Ensure that sterile spare parts are available for all parts that are to be
implanted.
• Only use products that correspond to the requirements of the
EC Directive 90/385/EEC:
— BIOTRONIK implant and blind plugs
— BIOTRONIK leads and lead introducer
— BIOTRONIK programmer with approved cable and adapter accessories
— External multi channel ECG recorder
— External defibrillator and paddles or adhesive electrodes
Unpacking the implant
Proceed as follows:
WARNING
Inadequate therapy due to defective implant
If an unpacked implant is dropped on a hard surface during handling, electronic
parts could be damaged.
• Use a replacement implant.
• Send the damaged implant to BIOTRONIK.
1
Peel off the sealing paper of the outer plastic container at the marked
position in the direction indicated by the arrow.
The inner plastic container may not come into contact with persons
who have not sterilized their hands or gloves, nor with non-sterile
instruments.
2
Use the gripping tab on the inner plastic container to remove it from
the outer plastic container.
3
Peel off the sealing paper of the sterile inner plastic container at the
marked position in the direction indicated by the arrow.
Note:
The implant is disabled on delivery and can be implanted immediately after
unpacking without manual activation.