36
Parameters
Blanking periods
PMT protection
Parameter
Range of values
Standard
Far-field protection after V
s
100 ... (10) ... 220 ms
100 ms
Far-field protection after V
p
100 ... (10) ... 220 ms
150 ms
Ventricular blanking after A
p
30 ... (5) ... 100 ms
30 ms
Parameter
Range of values
Standard
PMT detection/termination
OFF; ON
ON
VA criterion
250 ... (10) ... 500 ms
350 ms