Cardiac Airbag Technical Manual
77
a
all
Figure 21. Status at ERI
Upon reaching ERI, the battery has enough energy left to
continue monitoring for at least three months and to deliver
minimum of six high energy shocks. The estimates associated
with duration of ERI assume the ICD is sensing an intrinsic sinus
rhythm at a rate of 70 bpm. After this period the device is at
EOS and should be explanted. Upon reaching EOS
tachyarrhythmia detection and therapy is disabled.