62
Cardiac Airbag Technical Manual
Testing for diaphragmatic stimulation
1. Select
10.0 V
in the Pulse Amplitude pop-up menu
2.
4.
eturn to the patient’s intrinsic
to test for diaphragmatic stimulation.
Press
Start Pacing
.
3. Check for diaphragmatic stimulation.
Press
Stop Pacing
to r
rhythm.
Connecting the ICD to the leads
5. Verify
VT/VF Detection
is
disabled
.
6. Hand off the pre-programmed ICD to the implanting
physician.
7. Disconnect the ventricular lead pins from the PK-44
cable and insert each lead pin into the appropriate
ICD header port.
Pace/Sense pin = P/S V {IS-1}
HV 1 Shock pin = HV 1 {distal coil}
HV 2 Shock pin = HV 2 {proximal coil}
8. Tighten each set-screw using a BIOTRONIK torque
wrench until a clicking sound is heard.
9. Insert the ICD into the pocket with etching facing up.