Cardiac Airbag Technical Manual
41
Stored IEGM
- The ICD can store up to 30 minutes of two
ams (IEGMs) including the history
events:
•
Terminations
py
Figure 5. Episode List
channel intracardiac electrogr
and prehistory of the following
•
Detection
•
Redetection
•
Delivered
Shocks
The ICD can store IEGMs for the following events prior to ERI:
•
3 spontaneous VF episodes treated with shock therapy
•
Non-sustained VF episodes without shock thera
•
10 VT monitoring zone episodes
•
Induced episodes while the programmer wand is over
the implanted ICD
Following ERI declaration, no further EGMs are stored in the
ICD. However, the episode counters continue to update until
EOS is declared. (
See Figure
6)