Cardiac Airbag Technical Manual
53
Refer to the appropriate lead system technical
ontai
manual.
ner
d using standard aseptic technique
5.3 Opening the Sterile C
The Cardiac Airbag ICDs are packaged in two plastic containers,
one within the other. Each is individually sealed and then
sterilized with ethylene oxide.
Due to the double packing, the outside of the inner container is
sterile and can be remove
and placed on the sterile field.
ST_01
Peel off the sealing paper of the outer
oral region. The device may be
sue or in the muscle
e etched side facing
container as indicated by the arrow.
Do not contaminate the inner tray.
Take out the inner sterile tray by
gripping the tab. Open the inner tray
by peeling the sealing paper as
indicated by the arrow.
5.4 Pocket Preparation
Using standard surgical technique, create a pocket for the
device in the patient’s pect
implanted either below the subcutaneous tis
tissue. The ICD should be implanted with th
up. The leads should be tunneled or surgically brought into the
device pocket. If lead tunneling is performed, re-evaluation of
the baseline lead signals, after tunneling is recommended.