
60
Cardiac Airbag Technical Manual
Measuring R-wave Amplitude
1. Select
Intracardiac Measurements
2. Press and hold the
Record Data
(
See Figure
12)
button to measure the R-wave amplitude. There is
an audible tone when the ICD has detected a
sensed R-wave.
3. In general, R-wave amplitudes should be greater
than 5 mV.
4. Press
to obtain a printout of the measurement.
Figure 12. R-Wave Measurement
Содержание Cardiac Airbag
Страница 54: ...48 Cardiac Airbag Technical Manual Figure 8 Parameter Window...
Страница 55: ...Cardiac Airbag Technical Manual 49...
Страница 57: ...Cardiac Airbag Technical Manual 51...
Страница 73: ...Cardiac Airbag Technical Manual 67...
Страница 85: ...Cardiac Airbag Technical Manual 79...
Страница 90: ...84 Cardiac Airbag Technical Manual...