If the temperature selected is lowered again the alarm will stop until the heating plates
start to heat up again and at this point it will set itself at 1ºC above the new temperature
selected.
•
There is an alarm for temperatures over 45ºC in the bag probe during the preheating
process or in the irrigation probe during the field filling, treatment and field washing
processes. (working with a set temperature of 43ºC or less)
•
There is an acoustic and visual high priority alarm if the system does not heat up.
•
If the user increases the set temperature to over 43ºC, the temperature alarm will
activate at 2ºC over the set temperature selected.
If the set temperature is lowered again, the alarm will stop until the heating plates start
to heat up again and at this point it will set itself at 2ºC above the new set temperature
selected.
•
There is an alarm for the disconnection or erroneous reading of the irrigation
temperature probe, during the field filling, treatment and field washing processes.
•
Alarm for the disconnection or erroneous reading of the bag temperature probe, during
the preheating process.