4. Click the icon “Para” and you can set the parameters. (Baseline Factor,
Peak Threshold, Calculate…etc.)
14. Click “Run” to start the
process.
During Calibration
Please check the Alignment Marker has been located in correct position.
Software will recognize two peaks from Alignment Marker signal.
DO NOT
use Size
Marker or DNA sample doing “Calibrate” in a cartridge.
Brief introduction of the Signal Chart:
Содержание Qsep1