29
5-1. Unpacking Cartridge:
Please follow the steps of “Unpacking guide” inside the cartridge box.
5-2. Add 1X Separation Buffer into
“S” well and add diH
2
O into others.
* Buffer height should be equal to the groove of wells.
5-3. Alignment Marker preparation:
Add ≥ 20 μl “Alignment Marker 20 bp & 1000 bp” (C109100) into the 0.2 ml tube
and then add 10
μl “Mineral Oil” on the top of Alignment Marker.
6.
Click “Change Buffer”.
7.
Place Alignment Marker 20 bp & 1000 bp C109100 (≥ 20 μl) at MA-1 position
and the buffer tray on the tray holder.
* Please make sure buffer tray have been placed correctly and fitted in the
holder.
* Spin down and make sure the Alignment marker is in the bottom of tube and
no air bubble appears.
8.
Hold the plate by your hand and press down the tube of Alignment Marker
tightly into the well by thumb.