23
8.
Prepare buffer and alignment marker in 0.2ml PCR tubes and allocate at the
corresponding position on the sample tray
*For 12-well sample tray, user needs to use 0.1ml tube (C104252) for sample
and size marker
*S/P/C well can only place 0.2ml PCR tubes
*Make sure no air bubble appears in each tube
9.
Click “Change Sample” or “Change Buffer”, the sample door will open
automatically.
10.
Slide the buffer tray into the instrument. Make sure sample tray is pushed to
the end
11.
Click “Park”, the sample tray moves into instrument
12.
Open the cartridge door by pressing the white button and insert cartridge
(guiding groove facing front).
*Please follow cartridge unpacking guide to unpack the cartridge before using