Cartridge information will be presented on the main window after cartridge latched.
12-1. New Cartridge Calibration:
New cartridge needs to be calibrated before using.
Please proceed with the following steps.
•
Click “Confirm”.
•
Click “HV Check”.
* The storage and transportation condition may influence the gel-matrix
and cause unstable current. Check current and see if it’s stable or not.
After HV Check, if current is still unstable, please repeat this step 2-3 times.
•
Click “Calibrate”.
For troubleshooting, please refer to cartridge unpacking guide
troubleshooting part.
13.
Click on the blank column and designate the sample locations, test method,
sample duration, runs and result name following steps 13-1, 13-2, 13-3 and 13-
4.