C7SM-1.01
27/79
Service Manual of Cardio 7
2) LCD screen is blank although the power is on
LCD screen is
blank although
the power is on
Broken 34Pin
shield cable?
Change the
34Pin shield
cable
Does the LCD cable
have a poor condition
of connection or short-
circuit?
Check the
condition of the
connection and
then change if
it's not good
Is the LCD
module
unsatisfactory?
Change the
LCD Module
Normal
operation
Use
No
Yes
Yes
Yes
No
Yes
B
A
B
Ask the
manufacturer
about it, then take
action
A
A
A
No
No
Содержание Cardio 7
Страница 1: ...Cardio 7 SERVICE MANUAL Version 1 01 www ebionet com...
Страница 10: ...C7SM 1 01 10 79 Service Manual of Cardio 7 Left side view Handle Right side view Patient cable connection port...
Страница 12: ...C7SM 1 01 12 79 Service Manual of Cardio 7 LCD display...
Страница 35: ...C7SM 1 01 35 79 Service Manual of Cardio 7...
Страница 36: ...C7SM 1 01 36 79 Service Manual of Cardio 7 Section 4 Block Diagram Description 4 1 Circuit Block Diagram...
Страница 47: ...C7SM 1 01 47 79 Service Manual of Cardio 7 5 2 Internal Cabling...
Страница 48: ...C7SM 1 01 48 79 Service Manual of Cardio 7 A Lcd Assembly B 40Pin FFC Cable C New Battery Cable Assembly...
Страница 49: ...C7SM 1 01 49 79 Service Manual of Cardio 7 E QN TPH Cable F Motor Assembly G Power Switch Cable Assembly...
Страница 52: ...C7SM 1 01 52 79 Service Manual of Cardio 7 Bottom Cover Assembly Top Cover Assembly...
Страница 58: ...C7SM 1 01 58 79 Service Manual of Cardio 7 b Remove the 7 screws from Inside of Top cover...
Страница 68: ...C7SM 1 01 68 79 Service Manual of Cardio 7 6 2 Accessory...
Страница 69: ...C7SM 1 01 69 79 Service Manual of Cardio 7...
Страница 72: ...C7SM 1 01 72 79 Service Manual of Cardio 7 8 2 UI Board Top Bottom...
Страница 73: ...C7SM 1 01 73 79 Service Manual of Cardio 7 8 3 CPU Board Top Bottom...