Service Manual of Cardio 7
1-4. System Installation
Installation precautions
When installing cardio7 system, take the following precautions.
-
The equipment should be operated within 5
∼
40
℃
temperature and
10
∼
90
%
humidity.
-
Check the power cord connection and probe should be handled with care.
-
Do not plug several power cords in one electric outlet.
-
Main unit should be installed on an even, level place.
-
Check that the equipment is properly grounded to avoid noise.
-
The entire system settings are stored on the internal memory even when system
power is turned off and on again
-
Avoid excessive shock or vibration that could result in equipment damage.
-
Avoid using in the vicinity of dust and flammable material.
Power cable connection
The equipment operates when the power cable end connect into the Cardio 7 power cord.
Replace the fuse of the equipment if power is not turned on despite that power cord is
connected and battery power is in normal condition. Contact service dept. of Bionet Co.,
Ltd. if the equipment is not operated after replacing fuse.
Caution
For the continuous protection against the fire hazard, the fuse must be changed to
the same type and rate of fuse by technician (T5A 250V).
For the protection of environment, do NOT throw wasted fuse away in a dangerous
place for the example fire hazard
Patient cable connection
-
Connect the patient cable into the connection port at the rear of the equipment.
-
Connect the limb electrodes to the N, F, R, L lead and chest electrodes to C1, C2, C3,
C4, C5, and C6 of the patient cable.
Recording paper installation
-
Printer door open if printer release button at the side of Cardio 7 is pushed in the right.
Place the recording paper with the signal-recording side up and push back to close.
Содержание Cardio 7
Страница 1: ...Cardio 7 SERVICE MANUAL Version 1 01 www ebionet com...
Страница 10: ...C7SM 1 01 10 79 Service Manual of Cardio 7 Left side view Handle Right side view Patient cable connection port...
Страница 12: ...C7SM 1 01 12 79 Service Manual of Cardio 7 LCD display...
Страница 35: ...C7SM 1 01 35 79 Service Manual of Cardio 7...
Страница 36: ...C7SM 1 01 36 79 Service Manual of Cardio 7 Section 4 Block Diagram Description 4 1 Circuit Block Diagram...
Страница 47: ...C7SM 1 01 47 79 Service Manual of Cardio 7 5 2 Internal Cabling...
Страница 48: ...C7SM 1 01 48 79 Service Manual of Cardio 7 A Lcd Assembly B 40Pin FFC Cable C New Battery Cable Assembly...
Страница 49: ...C7SM 1 01 49 79 Service Manual of Cardio 7 E QN TPH Cable F Motor Assembly G Power Switch Cable Assembly...
Страница 52: ...C7SM 1 01 52 79 Service Manual of Cardio 7 Bottom Cover Assembly Top Cover Assembly...
Страница 58: ...C7SM 1 01 58 79 Service Manual of Cardio 7 b Remove the 7 screws from Inside of Top cover...
Страница 68: ...C7SM 1 01 68 79 Service Manual of Cardio 7 6 2 Accessory...
Страница 69: ...C7SM 1 01 69 79 Service Manual of Cardio 7...
Страница 72: ...C7SM 1 01 72 79 Service Manual of Cardio 7 8 2 UI Board Top Bottom...
Страница 73: ...C7SM 1 01 73 79 Service Manual of Cardio 7 8 3 CPU Board Top Bottom...