C7SM-1.01
23/79
Service Manual of Cardio 7
2-3. Diagnosis
Problem
Resolution
1. The diagnosis shows LVH.
1. If disease is diagnosed, abnormal ECG
always appears.
2. LVH occurs when the sum of one of V1 or
V2 and one of V5 or V6 is 3.5 or higher.
3. Upgrade to the latest version of the
software.
2. Diagnosis shows MI.
1. If MI is diagnosed, it is likely that the print
button was pressed before the wave
patterns are stabilized.
2. Check HR before printout and print it
again.
3. Upgrade to the latest version of the
software.
Содержание Cardio 7
Страница 1: ...Cardio 7 SERVICE MANUAL Version 1 01 www ebionet com...
Страница 10: ...C7SM 1 01 10 79 Service Manual of Cardio 7 Left side view Handle Right side view Patient cable connection port...
Страница 12: ...C7SM 1 01 12 79 Service Manual of Cardio 7 LCD display...
Страница 35: ...C7SM 1 01 35 79 Service Manual of Cardio 7...
Страница 36: ...C7SM 1 01 36 79 Service Manual of Cardio 7 Section 4 Block Diagram Description 4 1 Circuit Block Diagram...
Страница 47: ...C7SM 1 01 47 79 Service Manual of Cardio 7 5 2 Internal Cabling...
Страница 48: ...C7SM 1 01 48 79 Service Manual of Cardio 7 A Lcd Assembly B 40Pin FFC Cable C New Battery Cable Assembly...
Страница 49: ...C7SM 1 01 49 79 Service Manual of Cardio 7 E QN TPH Cable F Motor Assembly G Power Switch Cable Assembly...
Страница 52: ...C7SM 1 01 52 79 Service Manual of Cardio 7 Bottom Cover Assembly Top Cover Assembly...
Страница 58: ...C7SM 1 01 58 79 Service Manual of Cardio 7 b Remove the 7 screws from Inside of Top cover...
Страница 68: ...C7SM 1 01 68 79 Service Manual of Cardio 7 6 2 Accessory...
Страница 69: ...C7SM 1 01 69 79 Service Manual of Cardio 7...
Страница 72: ...C7SM 1 01 72 79 Service Manual of Cardio 7 8 2 UI Board Top Bottom...
Страница 73: ...C7SM 1 01 73 79 Service Manual of Cardio 7 8 3 CPU Board Top Bottom...