41
Battery Replacement: Control Unit
The battery in the Control Unit is a rechargeable AAA battery. It should be replaced
approximately every two years.
To replace the Control Unit battery (AAA NiMH 1.2 V):
1. Using the Phillips screwdriver, remove the screw from the battery cover on the back
of the Control Unit. See Figure 8-2. The screw is under a small label. Gently peel off
one end of the label. Reapply the label after the battery is replaced.
Figure 8-2: Replacing the Control Unit Battery
2. Remove the battery cover.
3. Note the "+/-" orientation of the old battery and remove the old battery.
4. Insert the new rechargeable battery in the proper "+/-" orientation.
5. Slide the cover into place, and tighten the screw.
6. Fully charge the new battery before first use.
Remove the old battery, and properly dispose of it according to your local environmental
regulations.
Caution: Use of a non-rechargeable AAA battery can damage the L300 System.
Chapter 8 - Maintenance and Cleaning
Control Unit
Battery
Battery Cover
Screw
Содержание NESS L300
Страница 1: ...NESS L300 User s Guide...
Страница 7: ...VII User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...
Страница 15: ...8 User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...
Страница 31: ...24 User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...
Страница 45: ...38 User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...
Страница 59: ...52 User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...
Страница 65: ...58 User s Guide Figure 9 5 Intelli Sense Gait Sensor Blank Label for the System ID Number Blank Label...
Страница 71: ...64 User s Guide T h i s p a g e i n t e n t i o n a l l y l e f t b l a n k...