M A C H - 1
U S E R M A N U A L V 8 . 8
34
The optional
“Data File Filter”
allows you to limit the number of data points in a file for long-
lasting experiments
. Selecting the “Data File Filter” checkbox enables the “Config.” button
Figure 38:
“Data File Path” Specified and the “Data File Filter” Checkbox Selec
ted
Clicking the “Config.” button will open the “Data File Filter” window.
Figure 39:
“Data File Filter” Window
The window provides the option to save
data when either the “Elapsed time (sec)”, the
“Displacement” (in mm)
,
or the “Loading” (in the load cell units)
is greater than or equal to a
user-defined value. The results will not be saved unless one of the specified conditions is met. To
deactivate a condition, set it
to “inf” (infinite) as the default “0” value is a co
ndition that is always
met.
The “Stage Axis” and “Load Cell Axis” can be chosen on multi
-axis systems.
Note:
If a file designated for saved results was chosen by mistake for one
function, click “Browse
Data Files” and “Cancel” to
remove it.
Note:
The data file filter cannot be used to manage images created by a camera feed (refer to
section “
Add-On for Camera Feed
”).
5.2.15
A Note on Acceleration
For all functions requiring a steady speed displacement (i.e. “Move Absolute / Relative”, “Ramp
-
Release” and “Stress Relaxation”) and for “Manual Controls” of the stage, Mach
-1 Motion
automatically sets the acceleration and deceleration at 20 times the requested velocity, up to the
maximal acceleration
. For the “Find Contact” function, the acceleration is set to 1 time
s the set
velocity.