M A C H - 1
U S E R M A N U A L V 8 . 8
16
the middle of the Calibration Holder (Figure 13.3 black arrow), while for a multiaxial load cell, the
calibration weight must be placed into either of the sides (Figure 13.3 red arrows). The system will
acquire the weight. A window (as shown in Figure 13.4) will appear. Remove the calibration
weight from the load cell and click “OK”. The system will acquire the absence of a weight.
The
load cell window (Figure 11) will now display the updated calibration information, which is
calculated automatically.
5 . 1 . 3 . 2
L O W - P A S S F I L T E R
Selecting
“Low
-
pass filter” reduc
es the noise in the load signal. It consists of a regular
Butterworth filter with user definable
“Order” and “Cutoff frequency,
Hz”.
Keep in mind that a
higher order will result in more phase shift of the signal.
5 . 1 . 3 . 3
A N A L O G I N P U T S ( A D D - O N F O R A N A L O G I N P U T S R E Q U I R E D )
Two Analog Inputs can be added to the Mach-
1™
Micromechanical Tester (see Figure 14). Use
the checkboxes to activate one or both analog inputs channels. For each channel the ground
reference setting can be set to
“Differential” or “
Single-Ended
”
and the voltage range set to
±10V, ±5V, ±1V or ±0.2V. Signals shall be connected to the dedicated SCC-68 box that is
provided to users with this software add-on. See the SCC-68 reference label inside the box and/or
the SCC-68 user guide for the proper location of the inputs terminals.
Figure 14
. “
Analog Inputs
” Window
5.1.4
User Manual
Use this feature to access a PDF file of this User Manual. Adobe Reader is required. Visit
5.1.5
About
Use this feature for information on which software version you have and for other information
related to your Mach-
1™
Micromechanical Tester.