26
(RAMP ON, RAMP OFF, ON TIME, OFF TIME and DELAY) in order
to elicit repeated muscle contractions and rest periods.
In the RECIPR mode Channel 1 and Channel 2 cycles alternate.
Therefore, Channel 1 and Channel 2 timing options can be set
completely independent of one another.
9.6 Setting the Reciprocal (RECIPR) Mode
Step 1:
Attach leadwires to Channel 1 (CH1) and if needed to
Channel 2 (CH2) (Refer to section 6.1)
Step 2:
Attach electrodes to the leadwires by following the
instructions on the electrode packaging.
Step 3: Attach electrodes to the body per the instructions given by
the prescribing practitioner.
Step 4:
Turn device on by pressing the "ON/OFF" Key (1)
Step 5:
Press the "Mode" Key (5)
until "RECIPR" appears on
the screen. The last settings used are displayed. The factory
settings are displayed if never used. To refer to factory settings
see section: 4.1 (Programmable/Adjustable Mode Starting
Parameters). To view the Timing options without changing them,
press the "Info" Key (7)
repeatedly to scroll through the
settings. To use the current settings proceed to Step 9. To change
the settings, proceed to Step 6.
CH 1
CH 2
ON TIME
OFF TIME