Note: Use of the Constant Mode or a Pulse Rate greater than 60Hz
with a symmetrical biphasic square waveform should be used for
the relaxation of muscle spasms only. Use of constant stimulation
and/or rates above the normal physiological range (e.g., greater
than 60Hz) could lead to rapid onset of muscle fatigue, making the
device less effective in producing repeated forceful muscle
contractions.
9.2 Setting the CONSTANT (CONST) Mode
Step 1:
Attach leadwires to Channel 1 (CH1) and if needed to
Channel 2 (CH2) (Refer to section 6.1 Connect electrodes and lead
wires)
Step 2:
Attach electrodes to the leadwires by following the
instructions on the electrode packaging.
Step 3:
Attach electrodes to the body per the instructions given
by the prescribing practitioner.
Step 4:
Turn device on by pressing the "ON/OFF" Key (1)
Step 5:
Press the "Mode" Key (5)
until "CONST" appears on
the screen. The last settings used are displayed. If the device was
never used, the factory settings are displayed. To refer to factory
settings see section: 4.1 (Programmable/ Adjustable Mode Starting
Parameters).
Step 6:
(Optio
nal) To change the Hz, μS or waveform; See
section 8.4 (Changing PR (Hz), PW (μS) and Waveform Values)
Step 7:
(Optional) To set the Patient Timer see section 8.4
(Patient Timer). If the timer is not set, the device will continue to
run until the "ON/OFF" key (1)
is pressed, or if the device is
idol (with no intensity settings) for 7 consecutive minutes.
Step 8:
To begin stimulation; increase or decrease the intensity of
each Channel by pressing the (2)
or (3)
buttons in the
"+" direction.