21
or Turn the device OFF and back ON by pressing the "ON/OFF"
Key (1)
Step 3:
To activate the lock hold down both the "ENTER" E (6)
key and the+ (4)
A key simultaneously for
approximately 3 seconds until the Clock symbol flashes on the LCD
screen. The Patient Lock Feature is now activated.
To Turn Off the Patient Lock Feature (unlock):
Press and hold both the ENTER E (6)
and the -(9)
simultaneously for approximately 3 seconds. Hold the buttons until
you see the Clock symbol on the LCD stop flashing and a double
digit number appears below the clock (This is the Patient
Compliance Meter read-out. See Section 8.7 Patient Compliance
Meter.) The Patient Lock Feature is now deactivated.
8.7 Patient Compliance Meter
The Patient Compliance Meter records usage time in hours
beginning from the first time the device amplitude is turned on
and provides stimulation to the user. It records and stores the
usage time in hours (up to 1099 hours). The Patient Compli ance
meter does not need to be "set" as it begins recording
automatically.
To view the amount of hours the patient has used the
device:
Step 1:
Turn on device by pressing the "ON/OFF" Key (1)
Step 2:
If any values are blinking on the LCD screen press the
"Enter Key" (6)
until all values are accepted and nothing is
blinking or Turn the device OFF and back ON by pressing the
"ON/OFF" Key (1)