25
Step 3:
Attach electrodes to the body per the instructions given
by the prescribing practitioner.
Step 4:
Turn device on by pressing the "ON/OFF" Key (1)
Step 5:
Press the "Mode" Key (5)
until "CYCLED" appears on
the screen. The last settings used are displayed. The factory
settings are displayed if never used. To refer to factory settings
see section: 4.1 (Programmable/Adjustable Mode Starting
Parameters). To view the Timing options without changing them,
press the "Info" Key (7)
repeatedly to scroll through the
settings. To use the current settings, proceed to Step 9. To
change the settings proceed to Step 6.
Step 6:
(Optional) To Change the Timing options refer to section
(s) 8.2-8.3 (Changing the Timing Options; RAMP TIME, ON TIME
OFF TIME & DELAY)
Step 7:
(Optional) To change the Hz, μ
S or waveform; See
section 8.4 (Changing PR (Hz), PW (US) and Waveform Values)
Step 8:
(Optional) To set the Patient Timer see section 8.5
(Patient Timer). If the timer is not set, the device will continue to
run until the "ON/OFF" key (1)
is pressed, or if the device is
idol (no intensity settings) for 7 consecutive minutes.
Step 9:
To begin stimulation; increase or decrease the intensity of
each Channel by pressing the (2)
or (3)
buttons in the
"+" direction.
The device will hold the custom settings until changed or until the
device is reset to factory settings. (See Section 8.8)
9.5 Reciprocal Stimulation (RECIPR)
Reciprocal Stimulation displayed as (RECIPR) is a mode which
delivers the set Pulse Rate (Hz) Pulse Width (US) and Waveform
(ASYM or SYM) in repeating cycles comprised of "timing options"
Содержание BioStim NMS2