31
Electri-
cal fast
transient/
burst IEC
61000-4-4
±2 kV for
power supply
lines
not
applicable
not applicable (for
INTERNALLY POWERED
ME EQUIPMENT)
Surge
IEC 61000-
4-5
± 1 kV line(s)
to line(s)
not
applicable
not applicable (for
INTERNALLY POWERED
ME EQUIPMENT)
Voltage dips
short in-
terruptions
and voltage
variations
on power
supply input
lines IEC
61000-4-11
<5% UT (>95%
dip in UT)
for0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT (30%
dip in UT) for
25 cycles
<5% UT (>95%
dip in UT)
for 5 sec
not
applicable
not applicable (For
INTERNALLY POWERED
ME EQUIPMENT
Power
frequency
(50Hz/60Hz)
magnetic
field IEC
61000-4-8
10V/m
10V/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test
level.
Guidance and manufacture’s declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment
specified below. The customer of the user of the device should assure
that it is used is such an environment.
Immunity
test
IEC 60601
Test level
Compli-
ance level
Electromagnetic
environment-guidance