30
12. ELECTROMAGNETIC COMPATIBILITY TABLES
Guidance and manufacture’s declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR11
Group 1
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11
Class B
The device is suitable for use in
all establishments including those
directly connected to the public
low-voltage power supply network
that supplies to buildings power used
for domestic purposes
Harmonic
emissions
lEC61000-3-2
Not applicable
Voltage fluctu
-
ations/ Flicker
emissions
lEC61000-3-3
Not applicable
Guidance and manufacture’s declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment
specified below. The customer of the user of the device should
assure that it is used is such an environment.
Immunity
test
IEC 60601 Test
level
Compli-
ance level
Electromagnetic
environment-guidance
Electrostatic
discharge
(ESD)
lEC61000-
4-2
±8kV direct
& indirect
contact;
±15kV air
discharge
±8kV
direct
& indirect
contact;
±15kV air
discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%