13
4
[-] button:
At treating mode, press the [-] button to decrease the intensity of the
CH1; At the key locking mode, press the [-] button to unlock the keys.
At setting mode, press the [-] button to decrease the pulse rate, pulse
width or treatment time.
5
[+] button:
At standby or treating mode, press the [+] button to increase the CH1
or CH2; At setting mode, press the [+] button to increase the pulse
rate, pulse width or treatment time.
6
[-] button:
At treating mode, press the [-] button to decrease the CH2
At the key locking mode, press the [-] button to unlock the keys.
At setting mode, press the [-] button to decrease the pulse rate, pulse
width or treatment time.
7
[P] button:
At standby mode, press the [P] button to select the treatment program.
At standby mode, press and hold [P] button to enter the setting mode.
8
Battery cover
9
Output socket
4. SPECIFICATION
Technical information
Device name
Combo Electrotherapy Device
Model/type
BioMed® Revived II
Power sources
4.5V D.C., 3x AAA batteries
Output channel
Dual channel
Waveform
Bi-phase square-wave pulse
Output current
Max. 120mA (at 500ohm load)
Output intensity 0 to 40 levels, adjustable
Treatment mode TENS, EMS and MASSAGE mode
Operating
condition
5° C to 40° C with a relative humidity of 15%-93%,
atmospheric pressure from 700 hPa to 1060 hPa
Storage condition -10° C to 55° C with relative humidity of 10%-95%,
atmospheric pressure from 700 hPa to 1060 hPa