9
Maintenance
NOTE:
To ensure the longevity of the battery power, only BIOLASE-supplied batteries are
recommended as replacements (BIOLASE p/n
2400203
); these are industrial-grade
batteries which under normal use have a longer life than conventional AAA batteries.
Figure 9.8 pairing button (internal view)
Figure 9.7 battery cover screws
9.5 CHANGING THE WIRELESS FOOTSWITCH BATTERIES
The wireless Footswitch is powered by two AAA batteries. When the batteries are low, a warning
message will appear on the display screen indicating that the batteries need to be replaced. To replace
the batteries, remove the 3 screws on the underside of the Footswitch to remove the battery cover (Figure
9.7); take out the old batteries and install the new ones, replacing the cover when done. Discard the used
batteries as regulated; do not th
r
ow them in a trash bin.
Do not press/push/touch the pairing button (Figure 9.8) while changing the batter
ies
, as this will disrupt
the pairing of the laser console and Footswitch.
Replacing the batteries may disrupt the pairing of the laser console and Footswitch. If you find the
wireless communication has been interrupted, reestablish pairing by following the instructions provided in
Section 7.