Figure 6.1: Power Switch, DC
Power Input, Remote Interlock
Figure 6.2:
DC
Power Supply
Module
17
6
Safety with the
Epic Hygiene
6.3 SAFETY FEATURES
ENERGY MONITOR
The energy monitor measures and verifies power output.
Power deviations of more than ± 20% from the selected
value will cause the display to show the error message:
“LASER CURRENT HIGH/LOW”.
The laser console will not operate until the system first clears
the error and then goes into
Ready
mode. If the error
message persists, please contact BIOLASE Service at
1-800-321-6717
or your BIOLASE-authorized service
representative.
SYSTEM MONITOR
The system monitors the emergency stop switch, remote key,
wireless Footswitch connection, and output power. An error in
any
one of these will stop the system. The text display will
indicate the
type of error. Operation cannot resume until the
error is cleared.
POWER SWITCH
The laser console can be switched ON ( I ) or OFF (O) using
the power switch on the back of the console. Turn the power
switch OFF (O) only when the system will not be in use for a
long period of time.
ACCESS KEY CODE
The Access Key Code (
888
) prevents unauthorized use of
the system. It is activated every time the system is turned ON
with the power switch.
Placing the laser in
Sleep
mode by pressing and holding the
Control button on the front panel of the console does not re-set
the access key code.
CONTROL BUTTON
Once the power switch is set to the ON ( I ) position, enter the
access key code. After setting the desired parameters for a
procedure, press the CONTROL button on the control panel to
enter into
Ready
mode. The aiming beam will illuminate to
indicate that the system is ready for use.
Epic Hygiene
User Manual
Содержание Epic
Страница 2: ...PAGE LEFT BLANK ...
Страница 51: ...Appendix F Electromagnetic Compatibility 51 Epic Hygiene User Manual ...
Страница 52: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 52 ...
Страница 53: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 53 ...
Страница 55: ...PAGE LEFT BLANK ...