1. Take the Handpiece body in one hand and the shaft in the other,
2. Pushing the two buttons on the fiber shaft,
3. Gently pull the Handpiece by the collar to separate it from the fiber shaft, without twisting.
7.4 SINGLE-USE TIPS
The Tips are single-use accessories
; a
400μm
size, available
in different lengths
, is recommended for
optimum results
(see Appendix A).
Figure7.5: Disconnecting the Handpiece from the Fiber Optic Cable
Figure 7.6: Insert the fiber Tip into the Handpiece (
only when the Handpiece is connected to
the fiber
) and twist clockwise until snug
CAUTION:
Tips are single-use only to avoid cross-contamination and are designed to
withstand only a single sterilization cycle; they must be disposed of after use in a
biohazard medical waste sharps container. Always visually inspect the Tip prior to use to
make sure it is free of debris or damage.
21
7
Installation and Set-Up
To connect the Tip, remove the protective cap from the end of the fiber shaft, connect the Handpiece
to the Fiber, then insert the tip firmly into the distal end of the Handpiece as far as it will go and
tighten by turning clockwise (Figure 7.6). Bend the metal cannula according the specific procedure
requirements (Figure 7.9).
Remove the fiber Tip by twisting the Tip counterclockwise (Figure 7.7).
Figure 7.4: Handpiece Assembly fully assembled
To disconnect the Handpiece from the Fiber Optic Cable (Figure 7.5):
Содержание Epic
Страница 2: ...PAGE LEFT BLANK ...
Страница 51: ...Appendix F Electromagnetic Compatibility 51 Epic Hygiene User Manual ...
Страница 52: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 52 ...
Страница 53: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 53 ...
Страница 55: ...PAGE LEFT BLANK ...