19
1. Clear the error message by pressing
. The display will show E
and .
2. Press the
and the blunger moves to the home-position. The
display will show E and TIP.
3. Remove the tip by pressing the tip ejection button
NOTE:
Repeated occurrence of error messages indicates an internal
error causing failure of the eLINE Dispenser to complete the execution
of the pipetting. The eLINE Dispenser therefore needs to be returned
to the nearest Sartorius Service Centre or your local Sartorius service
representative for repair.
Symptom
Possible cause
Solution
Display is blank and there is
no response when pressing
the buttons
Battery is fully
discharged
Recharge the battery
Leakage or pipetted volume
too small
Tip incorrectly attached
Attached new tip
Unsuitable tip
Attach firmly
Use original tips
Foreign particles between
the tip and tip connector
Clean the tip connector
and attach new tip
Dispenser out of the
established specifications
Unsuitable tip
Use original tips
Instrument damaged
Return to your Sartorius
service representative for
servicing
Tip ejector jammed or moves
erratically
Tip connector has been
contaminated
Clean the Tip connector
Tip connector mechanism
damaged
Return to your Sartorius
service representative for
servicing
Continuous error message
Instrument damaged
Return to your Sartorius
service representative for
servicing