3.3.4. Start Button
The start button triggers the aspiration and dispensing
operations according to the selected operating mode. Only a
soft touch is required to operate the button.
3.3.5. Tip Attachment and Ejection
Tip attachment
1. Push the tip into the tip connector. Make sure that the tip
is properly attached.
2. Confirm the tip size by double clicking the Start-button
according the instruction on the page 15.
Releasing the tip:
1. Press the Tip Ejection button. The tip is released
immediately.
NOTE:
After the dispensing cycle has been completed a small
residual amount of fluid may have been left in the tip. During
the tip ejection it is possible that a part of the remaining fluid
comes out of the tip. Therefore it is recommended that the tip
ejection is performed above a waste container.
NOTE:
When aspirating the liquid using the 25 or 50 ml tip,
let the tip stay in the liquid a few seconds after the plunger
has stopped moving. This way you will ensure that all liquid
has been aspirated into the tip.
NOTE:
It is possible to work with aqueous liquids and liquids
with higher viscosity. However, it is recommended always
to choose the tip with the smallest applicable volume in
conjunction with high viscosity liquids. Please note, the
specifications are given in accordance with ISO8655-5.
4. Charging of the Biohit eLINE
Dispenser Pro
The eLINE Dispenser should be charged using the original
eLINE Charging Stand, Charging Carousel or AC-adapter.
When the dispenser is properly connected to the charger, the
led indicator lights up and stays on as long as the dispenser is
connected to the charger.
When the dispenser requires charging, the battery symbol and
“lo bat”-text will appear in the display. You need to confirm
Led indicator
Battery symbol