Atomlab 500Plus (Dose Calibrator and Atomlab Wipe Test Counter
Combined)
Connecting the Detector and Chamber:
•
The Detector is included in its own box inside of the first carton. Open the box containing
the Detector and carefully remove it. Open the second carton and carefully remove the
Chamber. Place both the detector and the chamber in the area where they will reside.
•
Plug one end of RJ-12 Detector Cable 3 into the Interface Module as illustrated in the
circular inset at the upper left of Figure 7 making sure that the open port is used. Plug the
other end of the RJ-12 Cable into the Dose Calibrator/Detector as illustrated in the circular
inset at the top of Figure 7.
•
Plug one end of RJ-12 Chamber Cable 4 into the Dose Calibrator/Detector as illustrated in
the circular inset at the top of Figure 7and the other end into the Wipe Test Chamber as
illustrated at the bottom right of Figure 7.
Figure 7 Connecting the Detector and Wipe Test Chamber to the Interface Module
Note:
Do not apply power to the Atomlab until all of the connections have been
completed.