45
5061-049 REV 1.4
25-9574-75UM AE 05/2010
45
The concentration of the sample is displayed. Results shown as ---- indicate the
concentration is out of range.
Repeat step 7 for all samples.
Press
Esc
to return to the Applications Folder.
Press
to display available Options which are described below.
Step 8 (if using Standard mode)
Pipette on the reference sample and lower the sampling head. If Auto-Read is
off, press the 0A/100%T key. This will be used for all subsequent samples until
changed. If QA is switched on the sample will need to be replaced and the
0A/100% T key pressed again.
Press
to display the Run Standard Screen.
Run the standard by pressing
OR
Press cancel
Esc
to return to the Measure Screen.
Step 9
Clean the top and bottom plates, pipette on the sample and lower the sampling
head. If Auto-Read is off, press.
The concentration of the sample is displayed. Results shown as
---- indicate that the concentration is out of range.
Repeat step 8 for all samples.
Press
Esc
to return to the Applications Screen.
Press
to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
3. Toggles on/off, displaying a graph of wa/- 20 nm from selected
wavelength.
4. Return to Run Standard Screen.
7. Sample Number – add a prefix to the sample number and reset the
incrementing number to the desired value.
8. Save Method – use the left and right arrows to select a folder to store in
(Favourites/Methods 1–9), press the down arrow and enter name.
9. Auto-Print – toggles Auto-Print on/off.
Exit Options by pressing
Esc
, or wait.