If using a single cell holder, or a cell changer set to use as a single cell
holder, skip steps 18 through 19 and go straight to step 20.
Step 18
Press the batch measurement icon, then load the cell changer according
to the cell changer prompt. Then select the confirm icon.
Step 19
The acquired reference sample baseline will have be applied to all
subsequent sample measurements. The sample measurements can be
viewed by pressing the sample name test box and selecting the
appropriate sample from the list.
If using a cell changer, skip steps 20 through 21 and go straight to step 22.
Step 20
Insert the reference sample then take a reference measurement using the
reference measurement icon.
The acquired reference sample baseline will be applied to all subsequent
sample measurements.
Step 21
Replace the previous sample with a test sample then take a sample
measurement using the confirm key.
Repeat for all samples.
Step 22
Return to the home screen using the exit icon in the options menu
OR
use the additional options to save, print, and, or load previous
measurements (see the Additional Options section).