Ultrospec 7500 User Manual
•
41 56 2050 REV03
•
29
Step 14
Press the batch measurement icon, then load the cell changer according
to the cell changer prompt. Confirm when ready to take measurements.
Step 15
The acquired reference sample baseline will be applied to all subsequent
sample measurements. The sample measurements can be viewed by
pressing the sample name test box and selecting the appropriate sample
from the list.
If using a cell changer, skip steps 16 through 19 and go straight to step 20.
Step 16
Insert the reference sample then take a reference measurement using the
reference measurement icon.
The acquired reference sample baseline will be applied to all subsequent
sample measurements.
If “Absorbance”, “%Transmission” or “Concentration”
mode with the
“Predefined” factor method is selected, skip steps 17 through 18 and go
straight to step 19.
Step 17
If “Concentration” mode with the “Standards” factor method is selecte
d,
replace the previous sample with a standard sample of known
concentration then take a sample measurement using the sample
measurement icon.
Step 18
Enter the concentration value of the standard sample. Confirm the settings
using the confirm icon.